Before the analyzer: Why good laboratory results start with the sample

Veterinary medicine relies on sophisticated diagnostic technology, but technology cannot correct problems introduced before analysis happens.

AdobeStock

The chemistry analyzer can be perfectly calibrated; quality control procedures can be flawless. The veterinarian can order the right test, but if the sample sitting in front of the analyzer is hemolyzed, clotted, contaminated, underfilled, improperly stored, or collected in the wrong tube, none of that may matter.

Accurate laboratory results begin long before pressing "run."

Sample collection and handling can feel routine. We collect blood, fill tubes, make slides, spin samples, refrigerate specimens, and prepare shipments every day. That familiarity can make it easy to forget how much influence these simple steps have on the final result.

The period before laboratory analysis–the preanalytical phase–includes patient preparation, specimen collection, identification, handling, processing, storage, and transportation. Errors during this phase can alter results before the specimen ever reaches the analyzer. In one study performed at a U.S. veterinary teaching hospital, preanalytical errors predominated among identified laboratory errors.1

However, there is good news: Many of these preanalytical errors are preventable.

Start with the test, not the tube

Before collecting anything, ask yourself the question: What specimen does this test actually require? Whole blood, serum, plasma, urine, cytology samples, and microbiology specimens are not interchangeable. Neither are their containers.

For routine hematology, EDTA-anticoagulated whole blood is commonly used. However, collecting the sample is only the beginning. Storage time and temperature can affect hematologic measurements. Studies of canine EDTA samples have demonstrated changes in several CBC parameters as storage time increases, reinforcing the importance of timely analysis and appropriate storage.2,3

Chemistry testing commonly uses serum or heparinized plasma, depending on the laboratory and assay. Serum requires sufficient time for clot formation before centrifugation, whereas plasma can be separated without waiting for clot formation. Importantly, reference intervals and analytical methods may be based on a particular specimen type, making consistency important when monitoring a patient over time.

Coagulation testing raises the stakes even further. Sodium citrate tubes require a specific blood-to-anticoagulant ratio. An underfilled citrate tube is not simply a smaller sample–it can produce an inappropriate ratio and potentially inaccurate coagulation results.

The safest rule is straightforward: verify the laboratory's requirements before collecting an unfamiliar or specialized test.

Collection technique matters

A beautiful vein doesn't always produce a beautiful sample.

Traumatic venipuncture, excessive syringe suction, prolonged vessel occlusion, inappropriate needle size, forcing blood through a needle during transfer, or vigorous tube shaking can contribute to poor sample quality. Even variables such as feeding, stress, and physiologic rhythms may influence certain laboratory measurements and should be considered when interpreting results.4

Hemolysis is one of the most familiar examples. Ruptured erythrocytes release intracellular contents into the sample and can interfere with multiple laboratory measurements. Depending on the analyte and methodology, a hemolyzed specimen may produce results that do not accurately reflect the patient's actual circulating concentrations.

The solution is not complicated, but it requires consistency. Use appropriate equipment, obtain the sample as cleanly and non-traumatically as possible, allow evacuated tubes to fill appropriately, remove needles before transferring blood from syringes when indicated, and mix anticoagulated samples by gentle inversion rather than shaking.

A quick visual inspection before analysis or submission may identify hemolysis, lipemia, clot formation, inadequate volume, or other problems worth addressing before questionable results are received and documented.5

Illustration courtesy Jamie Morgan

Time and temperature are part of the equation

Once a blood sample leaves the patient, it remains biologically active. Cells continue to metabolize. Cellular components can leak into serum or plasma. Blood cell morphology changes with storage. Some analytes deteriorate rapidly, while others remain relatively stable. This means a specimen collected correctly can still become unsuitable due to delayed processing or improper storage.5

Refrigeration can help preserve an EDTA sample when immediate analysis is not possible, but storage-related morphologic changes still occur. Preparing fresh blood smears at the time of collection preserves a "snapshot" of cellular morphology that cannot be recreated several hours later.

Some tests have even stricter requirements. Ammonia is an example of an analyte requiring rapid processing and careful temperature control. Other specialized tests may require chilled, frozen, protected-from-light, or promptly separated specimens. Ultimately, storage requirements should be test specific.5

When less-than-ideal becomes unacceptable

All of us who have worked in veterinary medicine know collection is not easy, more so recollection. The patient may weigh two pounds. The unsocial cat has decided a second venipuncture is not happening. The dog may be hypotensive. The horse may be fractious. The owner may be waiting. So, what happens when the sample is less than perfect?

The answer depends on how the imperfection could affect the test we want to run. A mildly hemolyzed sample may still provide useful information for some assays while making other results unreliable. A small clot in an EDTA tube may invalidate platelet and cell counts. An underfilled citrate tube may compromise coagulation testing by causing an incorrect anticoagulant ratio. A severely lipemic specimen may interfere with some analytical methodologies. Delayed separation of serum or plasma may create changes that mimic or obscure an actual disease process going on.

Recognizing these limitations is part of good laboratory practice. When recollection is possible, and the sample defect could significantly influence the result, a fresh specimen is usually preferable to placing clinical confidence in questionable data. When recollection is not practical, document the sample quality, communicate with the veterinarian or laboratory, and interpret the results with the limitation in mind.5

The patient behind the sample

One of the easiest preanalytical errors to prevent has nothing to do with needles or blood tubes: it is proper identification. Every tube, slide, urine container, and other specimen container should identify the patient and collection site, particularly when multiple samples are collected. Submission information should accurately match the specimen.5

The same principle applies to documentation. Collection time, specimen type, medications, fasting status, collection site, and pertinent clinical history are all very important anytime sample testing occurs. When preanalytical factors cannot be completely controlled, documenting them can be important for accurate interpretation.4

Red flags before you hit 'run'

Before placing that sample on the analyzer–or sending it out the door–take 30 seconds and ask:

  • Is it the right patient? Confirm that the patient identification on the sample matches the medical record and laboratory request.
  • Do you have the right sample and the right tube? Verify the specimen type and collection container are appropriate for the requested test.
  • Do you have enough sample? Check minimum-volume requirements. Pay particular attention to anticoagulant tubes in which the blood-to-additive ratio matters.
  • Any clots? Inspect anticoagulated whole blood carefully. Even a small clot can affect cell counts, particularly platelet results.
  • Hemolysis or lipemia? After centrifugation, inspect serum or plasma Significant hemolysis or lipemia may interfere with certain laboratory measurements.
  • Was it handled correctly? Consider when the sample was collected, when it was processed, how it was stored, and whether the requested test has special temperature or handling requirements.
  • Something just doesn't look right? Stop before running the sample. Check the test requirements, contact the reference laboratory, or recollect when appropriate.

Quality begins with the patient and sample

Veterinary medicine relies on sophisticated diagnostic technology, but technology cannot correct problems introduced before analysis happens. Each step in the process–from collecting supplies, restraining, venipuncture, filling and mixing tubes, preparing blood smears, centrifuging samples, separating serum or plasma, labeling samples, storing samples, and preparing them for transport–is an opportunity to either protect the integrity of the specimen or compromise it.

The overall goal is to understand why collection and handling requirements exist, recognize when a sample may have been compromised, and know when a questionable specimen should be recollected. When it comes to laboratory diagnostics, accuracy does not begin with the analyzer, it begins with the sample.


Jamie Morgan, MAEd, CVT, RVT, CPEP, FFCP, is a veterinary technologist certified in both Illinois and North Carolina. She has more than 20 years of teaching experience in AVMA-accredited veterinary technology programs.

References

  1. Christopher, M. M., & Young, K. M. (2020). Frequency and classification of errors in laboratory medicine at a veterinary teaching hospital in the United States. Veterinary Clinical Pathology, 49(2).
  2. Médaille, C., Briend-Marchal, A., & Braun, J. P. (2006). Stability of selected hematology variables in canine blood kept at room temperature in EDTA for 24 and 48 hours. Veterinary Clinical Pathology, 35(1), 18–23. https://doi.org/10.1111/j.1939-165X.2006.tb00083.x
  3. Furlanello, T., Tasca, S., Caldin, M., Carli, E., Patron, C., Tranquillo, V., & Lubas, G. (2006). Artifactual changes in canine blood following storage, detected using the ADVIA 120 hematology analyzer. Veterinary Clinical Pathology, 35(1).
  4. Braun, J. P., Bourges-Abella, N., Geffré, A., Concordet, D., & Trumel, C. (2015). The preanalytic phase in veterinary clinical pathology. Veterinary Clinical Pathology, 44(1), 8–25.
  5. American Society for Veterinary Clinical Pathology. (2019). Principles of quality assurance and standards for veterinary clinical pathology (Version 3.0). Quality Assurance and Laboratory Standards Committee.